MICRO-SEPAROMETER MARK V DELUXE OPERATION MANUAL
13.0
FILTER SEPARATOR MONITOR TEST
13.1
The DELUXE MICRO-SEPAROMETER can be utilized to evaluate the condition of filter separators. A
proportionate sample is passed through a side-stream sensor to compare with the main-stream filter separator. The side-
stream element is then tested using the following method.
13.2
Prepare the DELUXE MICRO-SEPAROMETER as stated in section 7.0.
13.3
Momentarily depress the "ON" switch. The annunciator lamps located on switches A through G in the TEST
SELECT section will then be scanning for your selection. Depress switch "E". This will initiate the standard automatic
program for Filter Separator testing.
13.4
The annunciator lamp on the "START" switch will be illuminated which indicates the program can be started by
momentarily depressing this switch.
13.5
Momentarily depress the "START" switch. This will automatically operate the emulsifier for a period of 30 seconds
and raise the syringe drive to the up position.
13.6
After the emulsification period, remove the syringe from the emulsifier and insert the plunger with the help of the
wire aid. Remove the plug from the bottom, replace it with the F/S Monitor filter.
13.7
Momentarily depress "START" pushbutton. This will initiate the automatic program.
AUTOMATIC PROGRAM
Pulsed Tone
4 Seconds
Meter Adjust Period
10 Seconds
Syringe Drive Down
15 Seconds
Steady Tone
4 Seconds
Meter Read #1
10 Seconds
Wait
46 Seconds
Steady Tone
4 Seconds
Meter Read #2
10 Seconds
13.8
Upon completion of the F/S Monitor Program, the instrument will automatically reset.
COPYRIGHT 1984 EMCEE ELECTRONICS, INC.
REVISION DATE: FEBRUARY 12, 1987
PAGE 15