MICRO-SEPAROMETER MARK V DELUXE OPERATION MANUAL11.0CLEAR AND BRIGHT11.1A numeric value can be determined with the Micro-Separometer relative to sample haze. Using this procedure afuel sample is passed through a filter to remove water and particulate contamination. This sample provides a clear andbright reference which is compared to the original sample in the turbidimeter.11.2Prepare the DELUXE MICRO-SEPAROMETER as stated in section 7.0.11.3Momentarily depress the "ON" switch. The annunciator lamps located on switches A through G in the TESTSELECT section will then be scanning for your selection. Depress switch "C". This will initiate the standard automaticprogram for Clear and Bright Testing. The syringe drive mechanism will automatically move to the up position afterdepressing switch "C".11.4The annunciator lamp on the "START" switch will be illuminated which indicates the program can be started bymomentarily depressing this switch.11.5Remove a plunger from a 50 ml. syringe, insert a plug in the bottom and add approximately 50 ml. of fuel to betested. Insert the plunger with the help of the wire aid. Remove the plug from the bottom, replace it with a specialALUMICEL® COALESCER marked CLEAR & BRIGHT. Insert the syringe assembly into the syringe drive mechanism.11.6Momentarily depress the "START" pushbutton this will initiate the automatic program.AUTOMATIC PROGRAMSyringe Drive Down45 SecondsPulsed Tone4 SecondsMeter Adjust Period10 SecondsWait30 SecondsSteady Tone4 SecondsMeter Read10 Seconds11.7The automatic program starts by passing the test fuel through the ALUMICEL® COALESCER marked CLEAR &BRIGHT. Prepare the sample vial for fuel collection. The audio alert and the C/S indicator will indicate the proper timeto collect the last 15 ml. from the syringe.11.8Place the vial of filtered fuel into the turbidimeter and rotate to the marked position.COPYRIGHT 1984 EMCEE ELECTRONICS, INC.REVISION DATE: FEBRUARY 12, 1987PAGE 11
Integrated Publishing, Inc. - A (SDVOSB) Service Disabled Veteran Owned Small Business