MICRO-SEPAROMETER MARK V DELUXE OPERATION MANUAL
17.0
FIELD SERVICE
The DELUXE MICRO-SEPAROMETER has been calibrated and adjusted at the factory and should require very little
maintenance. This section, however, may assist in cases of minor difficulties that do not require factory service.
17.2
Syringe alert indication should a syringe alert light or light plus the audible alarm occur during testing, the following
steps should be taken: Repeat the test insuring the syringe is centered in the drive mechanism and not binding. Use
only a new syringe and plunger since the rubber on the plunger will swell causing increased drag. If possible, time the
total syringe down time with a stopwatch. If the time is correct 45±2 seconds (mode A) or 251 second (mode B) and the
alert is not repeated the instrument can be put back into service. If the syringe is within normal limits and the alert
system is still triggered the internal clocking system would be at fault and would require factory service. If the syringe
drive time is near or above the 47 second time limit it may be due to the syringe gear lubrication. The syringe drive
mechanism should be removed from the support panel and cleaned using a suitable solvent (Toluene or Kerosene) and
blow dry with an air pressure source. The gears and rack should then be lubricated with Molykote. After remounting the
syringe drive on the support panel, cycle the syringe drive several times and repeat the test. Should the drive time still
be out of limits factory service will be required.
COPYRIGHT 1984 EMCEE ELECTRONICS, INC.
REVISION DATE: FEBRUARY 12, 1987
PAGE 19