MICRO-SEPAROMETER MARK V DELUXE OPERATION MANUAL
11.9
When the pulsed tone starts the meter will automatically turn on. During this period adjust the display to 100 using
the meter adjust up or down switches.
11.10
Discard the fuel sample and place 15 ml. of unfiltered fuel into the sample vial. Insert sample vial into the
turbidimeter and rotate to the marked position.
11.11
When the steady tone starts the meter will automatically turn on. During this period record the display as the
Clear and Bright rating.