MICRO-SEPAROMETER MARK V DELUXE OPERATION MANUAL
9.9
When securely in place depress the "START" pushbutton. This will initiate the automatic program.
AUTOMATIC PROGRAM
Pulsed Tone
4 Seconds
Meter Adjust Period #1
10 Seconds
Emulsification
30 Seconds
Wait
30 Seconds
Pulsed Tone
4 Seconds
Meter Adjust Period #2
10 Seconds
Syringe Drive Down
45 Seconds
Wait
56 Seconds
Steady Tone
4 Seconds
Meter Read
10 Seconds
9.9.1 The automatic program starts with a read meter warning followed by a 10 second full scale adjustment period.
During this period adjust the display to 100 using the meter adjust up or down switches.
9.9.2 After the emulsification period, remove the syringe from the emulsifier and insert the plunger using the wire aid to
help expel the air. Remove the wire aid while maintaining the plunger at the 50 ml. position. Remove the plug from the
bottom, replace it with an ALUMICEL COALESCER, and place the whole assembly onto the syringe drive mechanism.
Attach ground lead between ground jack on side of syringe drive and ALUMICEL® COALESCER. Place a waste
container beneath the syringe to collect the unwanted fuel.
9.9.3 At the end of the second meter adjust period the syringe will start down. Remove and empty the sample vial and
prepare for sample collection. The audio alert and the C/S indicator will indicate the proper time to collect the last 15
milliliters from the syringe.
9.9.4 Place the sample vial in the turbidimeter and rotate to the marked position.
9.9.5 When the steady tone starts the meter will automatically turn on. An additional one second tone during the meter
on period will indicate the proper time to read the meter. Record the reading and report as Micro-Separometer rating
(MSEP rating).
COPYRIGHT 1984 EMCEE ELECTRONICS, INC.
REVISION DATE: FEBRUARY 12, 1987
PAGE 8