TM 11-6625-444-14-2
the readings indicated on the voltmeter are not
just A7-R40 until the reading displayed by the
within the tolerances specified, adjust the re-
voltmeter is within the tolerance specified by the
table.
(5) Proceed to the 20 V and 200 V ranges
Voltmeter Display
Dc Voltage
Adjustment
Tolerance
Calibrator (A1)
Range
Setting (Volts)
1.90000 0.00005
+1.90000
A7-R40
2V
+19.0000
A7-R39
19.0000 0.0005
20 V
+190.000
A7-R11
190.000 0.005
200 V