TM 11-6625-444-14-2
(1) Connect the equipment as shown in
back to the X10 position.
(2) Select the 200 V range and AC VOLTS
(15) Adjust A7-R83 so that the dc level of
mV.
(3) Select the 1 V range and 100 uV NULL
sensitivity on the differential voltmeter (A3) and
(16) There may be some interaction be-
set its dials to +0.0000.
tween the two adjustments. Repeat steps (10)
through (15) until both requirements are satis-
fied.
(4) Connect the dc differential voltmeter
and connect the HI terminal to A7-TP1 on the
(17) Set the differential voltmeter dials to
-.0003 and set its drum dial to 50 (i.e. -350 uV).
(5) Adjust the front panel AC NULL con-
(18) Adjust A7-R144 for zero null meter
trol on the voltmeter for reading of zero (10
indication (50 uV) on the differential voltmeter.
(19) Disconnect the differential voltmeter
f r o m A7-TP5 and connect the oscilloscope
(6) Disconnect the dc differential voltme-
ter HI test lead from A7-TP1 and connect it to
probe in its place. Set switch A7-S17 to the
NORM position.
(7) Adjust A7-R 144 for a reading of zero
(20) Change the oscilloscope volts/div
switch back to 5 mV/div.
(21) Adjust A7-R 113 slowly clockwise un-
NOTE
til a low frequency oscillation just appears on the
oscilloscope display. Rotate A7-R1 13 counter-
If the differential voltmeter indication
clockwise to the point where the oscillation just
tends to oscillate, adjust A7-R113
disappears.
slowly counterclockwise until the os-
cillation stops. Readjust A7-R144 for
a minimum reading on the differential
(22) Set switch A7-S17 back to the TEST
position and reconnect the differential voltmeter
test lead to A7-TP5. Disconnect the oscillo-
scope probe from the test point.
(8) Press the SLOW RESPONSE pushbut-
Adjust R148 for a
t o n to the "in" position.
(23) Set the differential voltmeter dials to
minimum reading on the differential voltmeter (
+.0002 (i.e. +200 uV). Adjust A7-R 144 for zero
10 uv). Allow a few minutes for the reading to
null meter indication ( 50 uV) on the differential
settle after initial adjustment.
If necessary,
repeat the adjustment. Release the SLOW RE-
SPONSE pushbutton back to the "out" position.
(24) Switch A7-S17 back to the NORM po-
sition.
The differential voltmeter null reading
(9) Switch A7-S17 to the TEST position.
should remain unchanged (50 uV).
(10) Set the volts/div switch on the oscil-
loscope to the 20 mV position and select dc
coupling. Set the time/div switch to 10 ms/div
c. Overall Zero Adjustments (A7-R94, R113,
and set the probe switch to the REF position.
R144, R148). Applies to units with serial number
Adjust the CH1 vertical position control to center
0 0 0 1 1 A thru 00039A, 00052A, 00055A and
the trace vertically.
00057A thru 00095A.
6-53