the retaining screws clockwise with a suitable
screwdriver until tight.
b. Preliminary Starting Procedure. Before us-
ing the TS-352B/U, carefully read the operating
instructions (para 2-4 through 2-7). For maxi-
mum accuracy in all measurements, use the range
that will produce a meter indication as close to
midscale as possible.
CAUTION
When measuring unknown voltage or
current values, start at the highest
range and reduce the range a step at a
time until midscale deflection is ob-
tained on the meter. Excessive voltage
or current values that cause greater
that full scale deflection can damage the
meter.
NOTE
Before using the TS-352B/U, check the
mechanical zero of the meter. If the
meter pointer is not exactly over the
zero line, reset by using the proper size
screwdriver to adjust the small screw
set into the glass of the meter face.
trols and indicators.
2-7