TM 55-4920-244-14
Malfunction
Probable cause
Corrective action.
1. Galvanometers (GALVO-1)
a. Defective switch SW-6.
a. Replace switch.
does not respond to tem-
b. Defective potentiometer.
b. Replace potentiometer
perature selector knob
adjustments with switch
c. Incorrect reference voltage to
c. See figure 4-25 and test
SW-6 in RANGE posi-
potentiometer.
following components of
tion (all applicable modes
cable and printed circuit
of operation).
board assembly. Part No.
BH1766; T2, CR8, CR9,
CR10, CR11, CR12, CR13,
C9, R14, R16, and R18.
Replace defective compon-
2. Tester inoperative in all
a. Defective power inlet con-
a. R e p l a c e power inlet connector
modes of operation.
nector.
3. POWER ON light inop-
erative.
b. Defective cable and printed
b. Replace cable and printed
circuit board assembly,
circuit board assembly
part no. BH1766.
c. Defective wiring or con-
c. Repair or replace wiring
nections.
4. Switch SW-1 in T/C posi-
a. Replace transformer (para.
tion; no light beam in
galvanometers (GALVO-
b. Redate cable and printed
b. Defective cable and printed
1).
circuit board assembly
circuit board assembly,
part no. BH1766.
c. Repair or replace wiring
c. Defective wiring or connec-
tions.
d. Replace switch .
d. Defective switch SW-1.
e. Repair or replace power lead.
e. Defective power lead cable to
cable and printed circuit
and (9)).
board assembly, part no.
BH1766.
a. Replace galvanometers (para.
a. Defective galvanometers.
5. Galvanometers (GALVO-1)
cannot be set to zero with
mechanical zero knob.
a. Replace potentiometer
6. Galvanometers (GALVO-1)
a. Defective potentiometer.
deflects erratically when
reading in RANGE po-
sition.
7. Heater probes will not heat;
a. Replace fuse and excite
a. Temporary power line over-
20 amp fuse blown.
voltage.
heater probes with TEMP
REGULATOR knob.
Continue operation if
fuse does not blow.
b. Heater cable connector de-
4-47).
fective or short in wiring.
c. Replace voltage regulator
c. Defective voltage regulator.
4-8
Change 1