TM 11-6625-539-15-2
(5) Place the 100-ohm, 1 percent, -watt
resistor between the E and B jacks on the front panel.
(6) Adjust resistor R30 so that meter Mlindi-
cates 100 ohms.
(7) Remove the 100-ohm resistor, and check to
see that the meter pointer indicates full scale. If the
meter pointer does not indicate full scale, repeat the
procedures in (2) through (6) above.
7-4. Alignment of Test
Set F.E.T. Position
a. Test Equipment Required. An ME-30(*) /U is
required.
b. Procedure.
(1) Disconnect the unit from the case by remov-
ing the four panel screws.
(2) Set the polarity switch to FWD PNP.
(3) Set the function switch to F.E.T. CAL.
(4) Place the ME-30(*)/U across the E and B
jacks on the front panel.
(5) Adjust CAL control R5 (fig. 7-2) and resis-
tor R23 (fig. 71 ) so that the meter pointer indicates
full scale when the ME-30(*)/U indicates 100 mv.