TM 11-6625-291-14/TO 33A1-7-23-1
2-10. MEASUREMENT OF OUTPUT POWER USING TS-585A/U. (CONT)
ACTION
ITEM
REMARKS
LOCATION
MEASUREMENT (CONT)
NOTE
If the characteristic impedance is unknown, see paragraph
2-12 to
find the characteristic impedance of
the circuit
being tested.
Set meter multiplier knob to 100.
3. Output Meter
Meter multiplier
TS-585AW
knob (4)
Set equipment under test on.
Meter multiplier
4.
Set meter multiplier knob to setting where
meter reading is slightly to right of center
knob (4)
scale.
More accurate readings are taken when
meter needle points near center of
scale.
NOTE
Read the power in milliwatts by: meter multiplier setting
times the meter reading. Read the power in decibels by:
meter multiplier setting plus the meter reading.
2-17