TM 11-6625-291-14/TO 33A1-7-23-1
2-9. OPERATING PROCEDURES.
The operating procedures listed below are given in this section:
Measurement
of output power using TS-585A/U
Measurement
of output power using TS-585B/U, TS-585C/U, or TS-585D/U
Measurement
of characteristic impedance using TS-585A/U
Measurement
of characteristic impedance using TS-585B/U, TS-585C/U, or TS-585D/U
2-10. MEASUREMENT OF OUTPUT POWER USING TS-585A/U.
This task covers:
Measurement
INITIAL SETUP
Tools
Personnel Required
One operator
None
Materials/Parts
Equipment Condition
Equipment under test off
None
ACTION
REMARKS
ITEM
LOCATION
MEASUREMENT
CAUTION
The output meter is not fused. Do not connect the output
meter to equipment rated more than 5 watts.
1. Output Meter
Test leads (1) and
Connect test leads to binding posts.
TS-585A/U
binding posts (2)
NOTE
If equipment under test is grounded, connect test lead from
negative binding post (marked--) to ground of equipment
under test.
Impedance control
2.
knob (3)
impedance setting of circuit being tested.
2-16