TM 11-6625-291-14/TO 33A1-7-23-1
2-11. MEASUREMENT OF OUTPUT POWER USING TS-585B/U, TS-585C/U, OR TS-585D/U.
This task covers:
Measurement
INITIAL SETUP
Tools
Personnel Required
None
One operator
Materials/Parts
Equipment Condition
None
Equipment under test off
ACTION
REMARKS
ITEM
LOCATION
MEASUREMENT
CAUTION
The output meter is not fused. Do not connect the output
meter to equipment rated more than 5 watts.
Test leads (1) and
1. Output Meter
Connect test leads to binding posts.
binding posts (2)
TS-585B/U
TS-585C/U, or
TS-585D/U
NOTE
If equipment under test is grounded, connect test lead from
binding post (marked G) to ground of equipment under test.
2.
Impedance control (3)
NOTE
If the characteristic impedance is unknown, see paragraph
2-13 to
find the characteristic impedance of
the circuit
being tested.
2-18