TM 55-4920-231-14
CAUTION
The case leak valve levers must be in
(3) When tests have been completed, stop
these positions and the safety valve
closed, to isolate the airspeed indicator
from the vacuum source to prevent dam-
age to the instrument.
(5) Make test hose connections (paragraph
CAUTION
38).
Do nut allow the altimeter reading to
go below -1000 f t or the rate-of-climb in-
dicator to exceed 6000 fpm as it will
CAUTION
damage the instruments.
Do not exceed +6000 fpm on rate-of-
a. 0 to 50,000 FT. Test.
climb indicator during tests as it will
(1) Perform the master instrument tests
damage the instrument.
(paragraph 39a).
(7) Turn vacuum INCREASE knob (17,
(2) Turn correction knob (22, figure 7) to
figure 7) to
open valve and
stabilize the
re-
provide a 29.92 reading on barometric scale (4)
on altimeter (3) and altimeter being bested.
(3) Place PRESSURE selector knob (10) in
(8) Calibrate the instrument by observing
OFF position and turn knobs (11 and 12) to
open valves.
to record the rate-of-climb readings on calibration
(4) Turn VACUUM selector knob (20) to
ALT & R.O.C. 0-50,000 FT position and turn
b. Rate-of-dive Test.
knobs (17 and 19) to close valves.
(1) Turn vacuum INCREASE knob (17,
CAUTION
figure 7) to close
valve and adjust vacuum DE-
The case leak valve levers must be in
CREASE knob (19) to provide the required
these positions, and the safety valve
closed, to isolate the airspeed indicator
(2) Calibrate the instrument by observing
from the vacuum source and prevent
damage to the instrument.
(5) Place case leak valve lever (3, figure
8) in
NOR position, case leak valve lever (4)
in CL position, and turn knob (14, figure 4)
to close safety valve (13).
CAUTION
Do not exceed +3000 fpm while testing
altimeters, nor +6000 fpm while testing
rate-of-climb indicators. Exceeding these
limits will damage the instruments.
NOTE
Tap altimeter lightly to remove friction before
(7) Turn vacuum INCREASE knob (17,
(8) Stabilize the required readings on al-
timeter (3, figure 7) by adjusting vacuum DE-
CREASE knob (19).
(9) Observe reading on instrument being
tested at each check point and record on calibra-
tion card (8).