Correction
Probable trouble
Symptom
Step
SHUNT potentiometer R33
open.
Check switches S5 and S19 and re-
GM-SIGNAL RANGE switch
place if necessary.
S5 or PRESS TO TEST P4
switch S19 defective.
Replace defective resistor or poten-
With GM-SIGNAL RANGE switch Resistor R40 or R42 in trans-
12
tiometer.
S 5 in F position, PERCENT
work or GM CENTERING
QUALITY meter M3 does not give
potentiometer R44 defective.
proper indication.
Check wiring and repair if defec-
With PRESS TO TEST P6 switch Wiring in grid circuit open
13
tive.
S21 depressed, indication on PER-
Replace defective resistor.
CENT QUALITY meter M3 drops Resistor R58 defective . . . . .
to zero.
Check wiring and repair if defect-
With tube tester adjusted for emis- Wiring in plate or meter cir-
14
ive.
cuit open.
sion test and PRESS TO TEST
P2 switch S17 depressed, PER-
CENT QUALITY meter M3 does
not indicate.
Replace defective resistor.
One of the current-limiting
resistors R19 through R25
in plate circuit defective.
Replace defective resistor or poten-
Shunt resistor R30, R31, R40,
tiometer.
R42, or GM CENTERING
potentiometer R44 defec-
tive.
Replace defective potentiometer.
With tube tester adjusted for vol- SHUNT potentiometer R33
15
tage-regulator test and PRESS
defective.
T O TEST P5 switch S20 de-
Replace defective potentiometer.
pressed, neither PECENT QUAL- SCREEN potentiometer R54
ITY meter M3 nor PLATE
or SIGNAL-V.R. potentio-
meter M5 indicate.
meter R46A defective.
Replace C4.
Screen supply filter capacitor
C4 short-circuited.
Check switch S4 and replace if
FUNCTION switch S4 defec-
necessary.
tive.
Replace defective potentiometer.
Operation of SIGNAL-V.R. con- SIGNAL-V.R.
16
potentiometer
trol R46 does not affect indica-
46B defective.
tion on PLATE meter M5.
Replace defective resistor.
Signal voltage divider resis-
tor R47, R49, or R50 defec-
tive.
Replace defective lamp.
With PRESS TO TEST PI switch FIL. CONT. SHORT lamp 12
17
S16 depressed and a tube known
defective.
Check switch S11 and replace if
to be good being tested, FIL.
necessary.
CONT. SHORT lamp 12 does not SHORT TEST switch S11 de-
light.
fective.
With tube tester adjusted for in- FUNCTION switch S4 defec-
Check switch S4 and replace if
18
necessary.
terelement leakage test, no deflec-
tive.
tion can be obtained on PLATE
Replace defective resistor.
meter M5 when there is leakage Multiplier resistor R38 defec-
between two or more elements of
tive.
tube under test.
former disconnected from the circuit.
AGO 1428A
22