TM 11-6625-2946-14/TM 08168A-14/1
Table 3-1. Operator/Organizational
Preventive Maintenance Checks and Services-Continued
W-Weekly
B--Before
M--Monthly
Q--Quarterly
Internal
Item
Item to reinspected procedure
Equipment will be reported not ready if:
no.
4
CONNECTOR%
Immediate corrective action is not possible.
Check front panel receptacles and accessory cable connectors
for bent or corroded pins or damaged locking mechanisms.
CABLES
5
Check cables for chafed, cracked or frayed insulation, and for
Replacement cables are not immediately available.
damaged connectors.
INTERIOR
6
Remove case (para 3-6) and inspect chassis and case interior.
Damage that causes equipment failure or reduces equipment
Remove any dust, moisture or corrosion observed. Inspect
assemblies and components for physical or heat damage.
7
PUBLICATIONS
Check to ensure that this technical manual complete.
usable, and current.
MODIFICATIONS
8
An URGENT MWO has not been applied,
Check DA Pam 310-7 and ensure that all applicable MWO's
have been applied. Marine Corps refer to SL1-2. URGENT
MWO'S must be applied immediately. Routine MWO'S
must be scheduled.
9
ENVIRONMENTAL PROTECTIVE FEATURES
Check case, gasket, and air valve for physical damage that
Damage is observed and equipment is to be used in a humid
would permit entry of moisture of foreign materials.
environment.
3-6. Removal
using TRICHLOROTRIFLUOROETHANE.
Prolonged breathing of vapor should be avoid-
a. Cover Removal. Loosen the two latch fasteners on
ed. The solvent should not be used near heat or
the front side of the case, raise cover, and slide cover to
open flame; the products of decomposition are
right until separated from case.
toxic and irritating. Since. TRICHLOROTRI-
b. Front Panel Removal. Using a Phillips screwdriver,
FLUOROETHANE dissolves natural oils,
loosen the ten captive screws on the perimeter of the
prolonged contact wit h skin should be avoided.
front panel and lift the front panel from the case.
When necessary, use gloves which the solvent
3-7. Cleaning
cannot penetrate. If the solvent is taken inter-
nally, consult a physician immediately.
Inspect the test set closely for deposits of dust, dirt, or
b. Remove grease, fungus, corrosion, and ground-in
other foreign substances. All surfaces, both external and
dirt from external surfaces with a cloth or soft brush
internal, should be clean and free of foreign matter.
dampened (not wet) with trichlorotrifluoroethane, item
a. Remove dust and loose dirt from the surface of the
case and front panel with a soft, lint-free cloth, item 2,
ponents, or irregular surfaces with a soft brush, item 1,
WARNING
Adequate ventilation should be provided while
Section
Ill.
TROUBLESHOOTING
3-8.
Troubleshooting
the replacement of physically damaged accessory items,
is restricted to the general support level.
All troubleshooting and repair of the test set, other than
3-2