MICRO-SEPAROMETER MARK V DELUXE OPERATION MANUAL
time is out of preset limits. The "C/S" indicator will light when 15 ml. of fuel sample is left for collection.
5.2.8 The TURBIDIMETER, located under the main control panel, consists of a well for placing the sample vial, a light
source and a photocell.
5.3
The small parts and supplies needed to carry out the test are shown in Figure 4 and consist of the following.
5.3.1 WIRE AID A piece of wire with a loop and hook twisted on one end, used to release air when the syringe plunger is
inserted. (Supplied with Instrument - Part No. 001-00-6928)
5.3.2 PIPETTE An automatic hand pipette for use with disposable plastic tip. (Supplied with Instrument Part No. 001-
00-5983)
5.3.3 DRIP PAN - A pan used to receive the waste fuel. (Supplied with Instrument - Part No. 001-00-7902)
5.3.4 CONTAINER - A clean container for water. (Not included)
5.3.5 SYRINGE (barrel and plunger) - A disposable 50 ml. plastic syringe. (Part of Six-Pack)
5.3.6 SYRINGE PLUG - A plug for the syringe. (Part of Six-Pack)
5.3.7 ALUMICEL® COALESCER A pre-calibrated aluminum, throwaway coalescer cell with a tapered end to fit the
syringe. (Part of Six-Pack)
5.3.8 VIALS - A glass vial pre-marked for proper turbidimeter positioning. (Part of Six-Pack)
5.4
SIX-PACK A new syringe, pipette tip, test sample vial, plug and ALUMICEL® COALESCER are used in each test.
These disposable parts are available from Emcee in a kit containing all the disposables for 6 tests. This kit is designed to
fit inside the Micro-Separometer. (Part No. 840.99-5944)
6.0
REAGENTS
6.1
WATER - Clean, surfactant-free, preferably distilled.
7.0
PREPARATION OF APPARATUS
7.1
If possible, locate the instrument in an area where the temperature is between 65 degrees F. and 85 degrees F.
COPYRIGHT 1984 EMCEE ELECTRONICS, INC.
REVISION DATE: FEBRUARY 12, 1987
PAGE 5