The RF Probe with its full-wave crystal detecting circuit
produces a true RMS response without turnover, or harmonic errors
for all voltage levels below 0.03 Volt, gradually changing to
peak-to-peak reading (calibrated in RMS) at approximately 1.0 Volt.
The probe has a shunt capacitance of approximately 2.5 pF at levels
of 0.3 volt or higher, increasing slightly at levels of 0.1 Volt or
less. (see Fig. 2) The shunt resistance component shown in Figure 1
is a variable factor, depending on the voltage and frequency.
A db scale in red on the panel meter and range switch
provides a convenient means of expressing relative voltage measure-
ments.
There is no zero control but instead a balancing control
allows residual indications due to noise to be minimized resulting
in excellent stability and ease of reading even on the most
sensitive range.
7