PROBE TERMINATIONS
a.
General: The probe terminations are factory sealed and their disassembly
is not recommended. If damaged, replace.
If a termination is suspected of being f a u l t y , first check that the center
b.
pin is making firm contact with the diode cartridge. An increase of mechanical re-
s i s t a n c e should be noted as the termination is within 1/16" of its final tightened
position.
Type 50-ohm termination adapter: Check the resistance value of the film re-
c.
sistor using the DC multimeter set to ohms. It should measure 50-ohm (49-ohm to
5 1 - o h m ) . If out of tolerance, replace.
Probe and diode cartridge assembly.
SWITCH CLEANING
A periodic lubrication and cleaning of the switch contact points is necessary for
prolonged accurate readings. They may be cleaned if necessary with the tip of a
soft cleaning cloth and a minimum of solvent without the disassembly of the
switch.