TM 11-6625-524-14-2
ME-30B/U
(3) Connect the second end of the CG-
UUT
409H/U to one side of the BNC Tee.
meter
range
(4) Connect one end of the second CG -
indication
min
max
selector
409H/U to the second side of the BNC Tee.
(volts )
position
(5) Connect the second end of the CG-
.003
.002
.00194
.00206
409H/U to the ME-30B/U.
.003
.003
.00291
.00309
(6) Connect the rf probe to the BNC tee
.01
.003
.00291
.00309
output .
.006
.01
.00582
.00618
.01
.01
.0097
.0103
AN/URM-127
the
c. Procedure. Adjust
.03
01
.0103
.0097
frequency to 500 KHz and the output amplitude
.02
.03
.0194
.0206
to the values shown below. The UUT readings
.03
.03
.0291
.0309
will be within tolerance.
.1
.03
.0291
.0309
.1
.06
.0582.
.0618
.1
.097
.1
.103
ME-30B/U
UUT
.1
.3
.097
.103
.2
.3
.194
.206
meter
range
.3
m in
max
.3
.291
.309
indication
selector
1
.3
.291
.309
(volts )
position
1
.6
.582
.618
.000285
1
.003
1
.97
.000315
.001
1.03
.006
.000570
3
1
.000630
.97
.001
1.03
2
.001
3
1.94
.001
.001050
2.06
.000950
.003
.001
3
.000970
2.91
3
.001030
3.09