f. R e l e a s e t h e P R E S S T O T E S T P 4 s w i t c h .
g. For twin-indicator tubes, operate the SCREEN and SUP-
PRESSOR SELECTORS switches to the positions specified in
t h e second line of the tube test data roll chart.
h. A g a i n d e p r e s s t h e P R E S S T O T E S T P 4 s w i t c h . I f t h e t u b e
is good, the indicator now should be open, or, in the case of twin-
indicator tubes, indicator No. 1 should be open and indicator No. 2
should be closed.
i. P e r f o r m a n e m i s s i o n t e s t o n t h e t r i o d e o f e l e c t r o n - r a y i n -
dicators with triode sections, such as 6U5 (par. 23). The control
settings for the emission test are specified in the third entry of
t h e same tube on the tube test data roll chart (par. 12).
b. O p e r a t e t h e F I L - S E L E C T O R S s w i t c h t o t h e p o s i t i o n s p e c i -
fied in the tube test data roll chart for the tube under test. All
other switches and controls should be in their safety positions
c. Insert the ballast tube into the appropriate socket.
d. Turn the ON-OFF power switch to the ON position.
e. Turn the SHORT TEST switch to the V position.
f. D e p r e s s a n d h o l d d o w n t h e P R E S S T O T E S T P 1 s w i t c h .
Operate the FIL+ SELECTORS switch to each of the positions
specified in the tube test data roll chart. The FIL.-CONT.
S H O R T lamp should glow in every position noted. If it does not,
discard the ballast tube.
g. R e p e a t t h e p r o c e d u r e s i n d i c a t e d i n b t h r o u g h f a b o v e f o r
each listing of the tube.
To check an indicator lamp or other lamps with miniature
b a y o n e t or screw-type bases, proceed as follows:
a. O P E R A T E T H E F I L - S E L E C T O R S s w i t c h t o p o s i t i o n 2 .
b. O p e r a t e t h e F I L + S E L E C T O R S s w i t c h t o p o s i t i o n 7 .
c. Operate the FILAMENT RANGE switch to the correct
voltage for the lamp under test. This voltage usually is marked
on the lamp base.
d . Operate the ON-OFF power switch to the ON position.
e. Adjust the FILAMENT fine control until the FILAMENT
VOLTS meter indicates the exact voltage specified for the lamp.