TM 11-6625-3060-10
Operating Procedures
Press the FULL SCALE 1 mv pushbutton switch. The dBm meter pointer should rest on the ZERO reference line. If it
does not connect the 50 ohm ADAPTER to the RF PROBE and use the ZERO control to set the pointer to the ZERO
reference line on the dBm meter. The negative sign (-) on the display should disappear.
VOLTAGE RANGES (mV) AND CREST FACTORS
VOLTAGE RANGES
1
3
10
30
100*
300*
1000*
3000*
(mV)
CREST FACTOR**
420
70
21
7
420
70
21
7
to
to
to
to
to
to
to
to
42
14
4.2
1.4
42
14
4.2
1.4
*With accessory 100:1 Voltage Divider
* *Maximum permissible ratio of peak to rms value of voltage.
2-9