TM 11-6625-291-14/TO 33A1-7-23-1
ACTION
LOCATION
ITEM
REMARKS
TESTING (CONT)
Output Meter
Voltmeter Reading Range (volts)
TS-585A/U
12.45 to 13.75
TS-585B/U
11.87 to 13.13
TS-585C/U
22.8 to 25.2
TS-585D/U
20.5 to 22.5
NOTE
In Output Meter TS-585A/U and Output Meter TS-585B/U, if
voltmeter does not read proper range, remove capacitor. If
voltmeter then reads properly, replace capacitor. If voltmeter
does not read properly after replacement of capacitor,
In Output Meter TS-585C/U and Output Meter TS-585D/U, if
voltmeter does not read proper range, replace auto-
6-31