TM 9-4935-282-34
Model
Maximum
Total
Current
Division Ratio
Division
Stability of
Temp. Coef.
No.
Input
Drawn At
500 V
1V
Accuracy
Division
of
Voltage
Max. Input
Out
Out
Accuracy
Division Ratio
Per Year
80A-1
1 KV
1M
1 ma
2:1
+0.015%
80A-2
2 KV
2M
4:1
80B-5
5 KV
5M
1 ma
10:1
5,000:1
+0.01%
0.005%
+0.005%/C
80B-10
10 KV
10 M
20:1
10,000:1
from 0 C to 60C
80C-15
15 KV
15 M
30:1
15,000:1
80C-20
20 KV
20 M
1 ma
40:1
20,000:1
+0.01%
+0.0005%
+0.0002%/C
80C-25
25 KV
25 M
50:1
25,000:1
from 0 C to 35C
80C-30
30 KV
30 M
60:1
30,000:1
MODELS
SIZE
INPUT
OUTPUT
CONNECTOR
CONNECTOR
80A
6-3/4" high
UG-560U with mating connector
Insulated binding posts on 3/4"
5-1/4" wide
supplied
centers
2-1/4" deep
80B
10-9/32" high
AN3102-18-165 with mating
Insulated binding posts on 3/4"
8" wide
connector supplied
centers for both outputs
10" deep
80C
13" high
Special 5" ceramic standoff with
Insulated binding post on 3/4"
9-3/4" wide
mating 8" polystyrene connector
centers for both outlets
16" deep
supplied
6-2. POTENTIOMETRIC RECORDER
The Fluke Model A-70 Self Balancing Potentiometric Recorder
is a special purpose high sensitivity recorder developed for
monitoring voltages measured by the
800 Series
Differential Voltmeters. In addition to providing a permanent
record of results, the recorder allows unattended monitoring of
various voltages from 0 to 500 volts for long periods of time. A
sample voltage which varies in direct proportion to the source
under measurement is obtained from the voltmeter null
detector by connecting the recorder to the output terminals at
the rear of the voltmeter. A gain adjustment control adjacent to
the voltmeter output terminals allows the sample voltage to be
adjusted such that full-scale meter deflection equals full-scale
recorder deflection. The large leakage resistance (500,000
megohms) between the recorder input posts and chassis is
such that the inherently high accuracy of the Fluke Differential
Voltmeter is generally not impaired in any way.
Electrical Span:
10 mv DC
Span Step Response Time:
Less than 0.5 seconds
Accuracy:
0.5% of span
Chart Speeds:
0.75, 1.5, 3.6, or
12 in. /min. or in./hr.
Dimensions:
Cabinet Model -
Cabinet Model--9-1/4" high, 8-1/4" wide, 15" deep.
Rack Model--Available on special request.
6-2