TM 11-6625-2946-14/TM 08168A-14/1
Table 5-7. Field Strength Measurement Circuit Performance Test
Control settings
Equipment
Step
Test
Test
Performance
no.
under test
procedure
equipment
standard
POWER: EXT a. Connect cable to output of the SG-1144(V)1/U and wrap five turns of
SG-1144(V)1/U
a. None
#20 insulated wire around the test set antenna.
FUNCTION:
Adjusted to measure
FSM
0.5 vrms, 30.050 MHz
b. Using a short (6" or less) clip lead, connect between ground on the
b. None
SC-I 144(V) l/U and the test set.
c. Adjust the output of the SG-1144(V)I /U for a level of 0.5 vrms and
c. Approx-
--serve reading on test set meter.
ob
imately 5.0 div.
5-17