TM 11-6625-291-14/T0 33A1-7-23-1
6-28. FINAL TESTING. (CONT)
ACTION
REMARKS
ITEM
LOCATION
TESTING OF METER MULTIPLIER CIRCUIT
Set to X10 X100 (1,000 ohms).
1. Output meter
knob (1)
Connect.
Meter test set (2)
2.
and binding posts (3)
Meter test set (2)
Set meter test set and meter multiplier knob
3.
meter multiplier
to each position listed in table below. Each
reading on meter should not vary more than
knob (4) and
0.25 dB.
meter (5)
NOTE
knob while performing this test.
Test Voltage from
Meter Multiplier
Meter Test Set TS-682/GSM-1
Control Knob Position
1.12
0.1
3.55
1.
11.2
10.
35.5
100.
6-50