TM 11-6625-3060-10
CHARACTERISTICS, CAPABILITIES AND FEATURES
The standard features of the ME526A/USM are:
1.
Programmable (logic level).
2.
Measures from 200 microvolts (uV)to3.0 volts from 10kilohertz (kHz) to 1.2 Ghz and 200millivolts(mV)to 300 V up
to 700 megahertz (MHz) using a 100 to 1.0 divider.
3.
True rms response 200 uV to 30 mV (to 3.0 V up to 700 Mhz using a 100 to 1.0 divider).
4.
Convenient pushbutton switches for ranging.
5.
BCD digital output plus dc analog output.
6.
7.
Overload protection to 400 volts dc (V dc), 10 volts (V ac).
8.
Vswr less than 1:1.15 up to 1.2 GHz.
EQUIPMENT DATA
Weight and Dimensions
Weight
4.1 kg (9 lbs. 3 oz.)
Length
292 mm (11.5 in.)
Width
211 mm (8.3 in.)
Height
132 mm (5.2 in.)
Input resistance: 500 k ohms to 10 k ohms (nominal)
(100 kHz to 100 MHz)
Accuracy:
1 mV to 3 V
Input Voltage
1% full scale (fs) plus
300 mV to 3 V
1% rdg
1% rdg.
3% rdg
10% rdg.
200 uV to 300 mV*'
2% rdg.
7% rdg
10
50
150
700
1.2
kHz
kHz
MHz
MHz
GHz
*Below 1 mV, add 1% fs
DATA PLATE
ME - 526A/USM
MFR 04901
SER NO.
CONTRACT DAAB07-82-G-H004
US
1-3