TM 11-6625-291-14/TO 33A1-7-23-1
ACTION
LOCATION
ITEM
REMARKS
TESTING
5. Binding post
Remove test set leads from binding posts.
Test set (3)
6.
Connect one test set lead to binding post
Test set (3)
not marked `G'.
7. Impedance network
Test set (3), short
Connect other test set lead to short segment
segment split
split ring of impedance network.
ring (4)
8.
Test set (3)
Read test set.
If test set does not show reading range of
bad. Refer to higher level of maintenance.
Output Meter
Test Set Reading Range (ohms)
TS-585A/U
30,351 to
31,168
TS-585B/U
31,383 to
32,017
TS-585C/U
17,523 to
17,877
TS-585D/U
17,186 to
17,533
6-37